ROCKLABS produce RSDs in a small bench top model and a standalone model. The Standalone RSD is designed to split out a representative portion of the sample – from 2% to 50%.
Specifications |
Motor power |
0.18kW, 3 phase |
Maximum input particle size |
10mm |
|
Width x Height x Depth |
1000 x 960 x 800mm |
|
Weight |
135kg |
|
Hopper capacity |
25kg |
|
Dust extraction |
1 extraction port available |
|
Options |
Options Sample bins: |
|
Warranty |
ROCKLABS Standalone RSD is sold with a 2000 operating hours or one year warranty (whichever comes first), covering all parts (excluding wear parts) and labour if used under laboratory conditions. |
|
Comments on taking |
Taking a representative split from a large sample is not as simple as many people believe. Rolling on a mat, multiple grabs or riffle splitting do not produce accurate results. A classic study* conducted at Bradford University demonstrated that a rotating sample divider provided the best results. |
|
* Critical evaluation of powder sample procedures. Allen T and Khan A. A, The chemical Engineer, May 1979, pp108 - 112 |
||
Standard deviation of samples % |
||
Dividing |
Cone and quartering |
6.810 |
Scoop sampling |
5.140 |
|
Riffle splitting |
1.010 |
|
Table sampling |
2.090 |
|
Rotating splitting |
0.125 |
|
Random variation |
0.076 |
|
Shipping Dimensions |
Width x Height x Depth |
1570 x 1160 x 920mm |
Weight |